Guangshu Electrical Technology Co., Ltd
new new

new

Home >  new > 

300W medical grade switching power supply with 24V output

Time:2025-12-03 Views:0

  I. Core Working Principle (Medical-Safety Focus)

  Medical-Grade Isolation & Regulation Architecture: Integrates a reinforced isolation AC-DC conversion stage (90-264VAC universal input, active PFC with PF≥0.95) and precision 24V DC output module, compliant with IEC 60601-1 (Edition 3.1) medical safety standards. Key design highlights:

  Isolation Level: 2 MOPP (Means of Patient Protection) or 4 MOPP (for critical care equipment) to prevent electric shock, with isolation voltage ≥4000VAC (input-output) and ≥1500VAC (input-ground).

  Leakage Current Control: <100μA (patient-accessible parts) and <500μA (operator-accessible parts)—critical for near-patient medical devices (e.g., monitors, infusion pumps).

  24V Output Stabilization: Uses synchronous rectification + feedback control to maintain ±1% output accuracy (even at 10%-100% variable load), avoiding interference with medical sensors.

  Medical-Scene Adaptation: Optimized for low noise (≤30dB) via fanless convection cooling (or low-speed PWM fans) to avoid disturbing patients; wide operating temperature (-20℃~60℃) compatible with hospital environments (e.g., cold storage rooms, warm wards); and anti-corrosion coating on PCBs to resist disinfectant exposure (e.g., ethanol, hydrogen peroxide).

  Reliability for Critical Care: Adopts “fault-safe” design—transient faults (e.g., voltage spikes) trigger immediate output shutdown (with visual/audio alerts) instead of unstable output, preventing harm to patients; supports optional 1+1 redundancy for life-support equipment (e.g., ventilators).

  II. Key Features (Medical-Grade Optimization)

  1. Safety Certifications & Isolation (Non-Negotiable for Medical Use)

  Mandatory Medical Certifications:

  IEC 60601-1:2018 (Edition 3.1) + regional variants (UL 60601-1, GB 9706.1-2020) — ensures compliance with global medical device safety norms.

  Isolation Classification: 2 MOPP (for general medical equipment, e.g., ultrasound machines) or 4 MOPP (for critical care, e.g., cardiac monitors).

  Leakage Current Limits: <100μA (patient leakage) per IEC 60601-1 Clause 8 — avoids micro-shocks to patients with compromised skin (e.g., burn victims).

  Electric Shock Protection:

  Double Insulation: Reinforced plastic enclosures (V0 flame rating) + internal insulation barriers to prevent direct/indirect contact with live parts.

  Over-Voltage Protection (OVP): 115%-130% of 24V (e.g., 27.6V) — shuts down output to avoid damaging medical device circuits (e.g., sensor modules).

  2. 24V Output Performance & Reliability

  Electrical Specifications:

  Output Voltage: 24V DC (fixed), ±1% accuracy (load 10%-100%) — critical for precise control of medical devices (e.g., infusion pump flow rates).

  Ripple & Noise: ≤50mVp-p (20Hz-20MHz) — minimizes interference with sensitive medical signals (e.g., ECG, EEG readings).

  Load Capacity: 300W (24V/12.5A) continuous, 10% peak overload (330W/13.75A for 10s) — handles transient loads (e.g., motor startup in surgical tools).

  Long-Term Reliability:

  MTBF (Mean Time Between Failures): >200,000 hours (per MIL-HDBK-217F at 25℃) — uses 105℃ long-life electrolytic capacitors and medical-grade MOSFETs.

  Environmental Resilience: -20℃~60℃ operating temperature, 10%-90% RH (non-condensing) — compatible with hospital wards, ICUs, and mobile medical units.

  3. Medical-Scene Friendly Design

  Low-Noise Operation:

  Fanless Cooling (for 200W max load): Convection cooling (no moving parts) ensures ≤25dB noise — ideal for patient-facing devices (e.g., bedside monitors).

  Low-Speed PWM Fan (for 300W full load): 800-2000 RPM (≤30dB) with dust-resistant bearings — avoids disturbing patients during long-term use.

  EMC Compliance:

  Emission: EN 55032 Class B, FCC Part 15 Class B — prevents interference with hospital IT networks and medical equipment.

  Immunity: EN 61000-4-2 (ESD), EN 61000-4-3 (RF), EN 61000-4-4 (EFT) — resists electromagnetic disturbances from defibrillators, MRI machines, etc.

  Installation Flexibility:

  Compact Form Factors: 150×100×35mm (bench-top) or 180×120×40mm (rack-mount) — fits inside medical device enclosures (e.g., dialysis machines) or hospital racks.

  Wiring: Touch-safe terminal blocks (IP2X) to prevent accidental contact during maintenance — compliant with IEC 61439-1.

  III. Recommended 300W Medical Grade 24V Power Supplies

  MedPro 300-24F (Fanless Near-Patient Grade):

  Core Highlights: IEC 60601-1 2 MOPP, <100μA patient leakage, fanless (≤25dB), 24V/12.5A, 85% efficiency, 3-year warranty.

  Ideal For: Bedside monitors, infusion pumps, portable ultrasound devices (patient-facing, low-noise requirements).

  Key Specs: 150×100×35mm, 500g, 90-264VAC input, ±1% output accuracy, -20℃~60℃ operating temp, IP21 (splash-proof for disinfectant).

  MedRedun 300-24R (Critical Care Redundant Grade):

  Core Highlights: IEC 60601-1 4 MOPP, 1+1 redundancy, <50μA patient leakage, 24V/12.5A (per module), 88% efficiency, 5-year warranty.

  Ideal For: Life-support equipment (ventilators, cardiac monitors), surgical robots (zero-downtime requirements).

  Key Specs: 180×120×40mm, 800g, 90-264VAC input, redundant failover <10ms, remote monitoring (RS485), -10℃~65℃ operating temp.

  MedCompact 300-24P (Portable Medical Grade):

  Core Highlights: IEC 60601-1 2 MOPP, <80μA patient leakage, lightweight (400g), 24V/12.5A, 82% efficiency, 2-year warranty.

  Ideal For: Portable defibrillators, mobile ECG machines, field medical units (small size, low-weight requirements).

  Key Specs: 140×90×30mm, 400g, 90-264VAC input, battery backup interface (12V Li-ion), IP20 (dust-proof).

  IV. Purchasing & Usage Guidelines

  1. Selection Criteria (By Medical Scenario)

  By Device Type:

  Near-Patient Devices (monitors/pumps): MedPro 300-24F (fanless, low leakage, low noise).

  Critical Care (ventilators): MedRedun 300-24R (4 MOPP, redundancy, remote monitoring).

  Portable Devices (field ECG): MedCompact 300-24P (small size, lightweight, battery backup).

  By Isolation Requirement:

  General Medical: 2 MOPP (MedPro/MedCompact) — sufficient for non-life-support.

  Critical Care: 4 MOPP (MedRedun) — mandatory for equipment with direct patient contact (e.g., dialysis catheters).

  By Environmental Exposure:

  Disinfectant-Prone Areas (wards): IP21 (MedPro) — resists ethanol/sodium hypochlorite splashes.

  Mobile Use (field): IP20 + wide temp (-20℃~60℃) (MedCompact) — handles harsh transport conditions.

  2. Installation & Maintenance Tips

  Pre-Installation Compliance:

  Verify Isolation: Use a hipot tester to confirm input-output isolation ≥4000VAC (per IEC 60601-1) — avoid using modules with degraded insulation.

  Leakage Current Test: Measure patient leakage current (must be <100μA) with a medical-grade leakage tester — critical before connecting to patient-facing parts.

  Routine Maintenance:

  Monthly: Inspect terminal blocks (tightness, corrosion) and clean fan filters (if equipped) — prevent overheating from dust buildup.

  Quarterly: Re-test leakage current and isolation resistance — degradation over time (e.g., PCB corrosion) can increase leakage.

  Annually: Replace electrolytic capacitors (even if functional) — medical-grade caps degrade after 5-7 years, risking output instability.

  3. Pitfall Avoidance (Medical-Specific Risks)

  Reject “Industrial-Grade Disguised as Medical”:

  Avoid models claiming “medical-grade” without IEC 60601-1 certification — industrial standards (UL 60950-1) do not meet leakage current or isolation requirements (e.g., industrial leakage <1mA vs. medical <100μA).

  Verify Certification Authenticity: Check the manufacturer’s IEC 60601-1 test report (not just a logo) — ensure it covers Edition 3.1 (latest standard, mandatory in most regions).

  Don’t Ignore EMC Interference:

  Medical devices (e.g., ECG machines) are sensitive to ripple noise — reject 24V models with >100mVp-p ripple (causes signal distortion).

  Test immunity to hospital EMI: Ensure compliance with EN 61000-4-4 (EFT) — electrical fast transients (from defibrillators) can trigger false shutdowns.

  Avoid Underestimating Redundancy for Critical Care:

  Life-support equipment (ventilators) requires 1+1 redundancy — a single power supply failure can be fatal; MedRedun 300-24R’s failover <10ms ensures uninterrupted operation.

Read recommendations:

5V2A Wall Plug Australian Standard

65W Switching power supply

200W desk Switching power supply

Power Adapter Warranty Policy